Fluke 52120A Transconductance Amplifier – Operational Used Unit

$0.00

Quantity
SKU: N/A Category: Brand:

Description

Additional Features:

  • Operating limits:
    • Output ranges: 2 A, 20 A, 120 A (100 A DC)
    • Input current range: 200 mA, 200 mA, 120 mA (100 mA DC)
    • Current gain: 10, 100, 1000
    • Input voltage range: 2 V, 2 V, 1.2 V (1.0 V DC)
    • Transconductance: 1,10,100 Siemens
    • Frequency: To 10 kHz
    • Maximum output compliance voltage: 4.5 Vrms (6.4 Vpeak)
    • Inductive drive capability:     1 mH load
    • Output isolation, current terminal to earth: 600 Vrms, 850 Vpeak, DC to 850 Hz
  • Performance specifications:
    • AC accuracy, closed loop with 6105A: To 140 ppm
    • DC accuracy, standalone: To 160 ppm
    • AC accuracy, standalone: To 350 ppm
    • Phase angle accuracy: 0.006°
    • Load dependent phase angle shift: <0.001° @ 60 Hz
  • General specifications:
    • Input line voltage range: 100 V to 240 V with up to ±10 % fluctuations
    • Transient overvoltage: Impulse withstand (overvoltage); Category II of IEC 60364-4-443
    • Frequency: 47 Hz to 63 Hz
    • Maximum consumption: < 1500 VA
    • Calibration Documentation: Report of calibration with data included; 17025 accredited report optional
    • Dimensions with feet (H x W x L): 192 mm x 432 mm x 645 mm (7.6 inches x 17 inches x 25.5 inches)
    • Dimensions without feet (H x W x L): 178 mm x 432 mm x 645 mm (7 inches x 17 inches x 25.5 inches)
    • Weight: 25 kg (54 lb.)
    • Design standards and compliance: Designed to EN 61010-1: 2010, CAN/CSA 22.2 No 61010.1-04, ANSI/UL 61010-1:2004, EN 61326-1:2006 CE marked, CSA listed
    • Operating temperature: 5 °C to 35 °C
    • Calibration temperature range: 16 °C to 30 °C
    • Warm up time: 1 hour
    • Safe operating max. relative humidity (non-condensing): <80 % 5 °C to 31 °C ramping linearly down to 50  % at 35 °C
    • Operating altitude: 0 m to 2,500 m
    • Shock and vibration: MIL-PRF-28800F class 3
    • Specification confidence level: 99%

Explore more from our collection.

Additional information

Calibrations

NIST Traceable, NIST Traceable with Full Data, ISO IEC 17025 Accredited

Reviews

There are no reviews yet.

Be the first to review “Fluke 52120A Transconductance Amplifier – Operational Used Unit”

Your email address will not be published. Required fields are marked *